UH LADCP Documentation

Navigation

This Page

WH300 annotated command fileΒΆ

;======================================================================
;                    M A S T E R . C M D
;                    doc: Tue Jun 15 11:46:07 2004
;                    dlm: Mon Oct 30 12:22:23 2006
;                    (c) 2004 ladcp@
;                    uE-Info: 42 0 NIL 0 0 72 2 2 8 NIL ofnI
;======================================================================

; This is the default master/downlooker command file

; NOTES:
;     - contains only commands that change factory defaults
;     - beam coordinates
;     - staggered single-ping ensembles every 1.5s/2.0s
;     - 25 8m cells --- reduce after determining the regional instrument

; HISTORY:
; Feb  1, 2006:
;     - released for CLIVAR P16N
; Oct 30, 2006:
;     - adapted for LADDER 1

CR1                   ; return to factory default

TC2                   ; ensembles per burst
LP1                   ; pings per ensemble
TB 00:00:03.50                ; time per burst
TE 00:00:01.50                ; time per ensemble
TP 00:00.00           ; time between pings

LN25                  ; number of depth cells
LS0800                        ; bin size [cm]
LF0                   ; blank after transmit [cm]

LV250                 ; ambiguity velocity [cm/s]


EZ0011101             ; Sensor source:
                      ;       - manual speed of sound (EC)
                      ;       - manual depth of transducer (ED = 0 [dm])
                      ;       - measured heading (EH)
                      ;       - measured pitch (EP)
                      ;       - measured roll (ER)
                      ;       - manual salinity (ES = 35 [psu])
                      ;       - measured temperature (ET)

EX00100                       ; coordinate transformation:
                      ;       - radial beam coordinates (2 bits)
                      ;       - use pitch/roll (not used for beam coords?)
                      ;       - no 3-beam solutions
                      ;       - no bin mapping

CF11101                       ; Flow control:
                      ;       - automatic ensemble cycling (next ens when ready)
                      ;       - automatic ping cycling (ping when ready)
                      ;       - binary data output
                      ;       - disable serial output
                      ;       - enable data recorder

CK                    ; keep params as user defaults (across power failures)
CS                    ; start pinging